Frequently Asked Questions (FAQs) regarding Assisted Blood Glucose Monitoring and Insulin Administration
 
On this Page
 
   
Background   The following FAQs summarize  inquiries from healthcare personnel received by CDC regarding best practices  for performance of assisted blood glucose monitoring and insulin administration,  including questions related to cleaning, disinfection, and storage of blood  glucose monitoring equipment.  These FAQs are not intended as a  comprehensive resource for all issues related to blood glucose monitoring, and  insulin administration and additional considerations may be necessary for  certain clinical situations or settings.  View more detailed information related to assisted blood glucose monitoring  and insulin administration. Visit CDC's Injection Safety website for additional  information regarding injection safety and CDC's Sharps Safety website information related to sharps safety and safe disposal of sharps in healthcare settings.   Healthcare personnel are also  encouraged to consult guidance provided by the Food and Drug Administration  (FDA) (links provided in responses below) as well as the manufacturers of the  devices (blood glucose meters, fingerstick/lancing devices, insulin pens) in  use at their facilities.  General      - What is the difference between “self-monitoring  of blood glucose” (SMBG) and “assisted monitoring of blood glucose” (AMBG)?     
With self-monitoring of blood glucose, individuals  perform all steps of monitoring for themselves.   With assisted monitoring of blood glucose, the same steps are followed  but testing is performed for an individual or multiple persons by someone else  (e.g., a caregiver or healthcare professional) [1, 2].  Assisted monitoring of blood glucose is  typically performed in healthcare settings such as clinics, hospitals, and  long-term care settings (e.g., skilled nursing facilities and assisted living  facilities).      Individuals who perform blood glucose monitoring  either for themselves or on others must be aware of basic safe practices to  protect against infection transmission.   These include the following infection control requirements:         - Fingerstick devices,  also called lancing devices, should never be shared, even with close  family and friends. This guidance includes both the lancet (i.e., the sharp  instrument that actually punctures the skin) and the pen-like device that  houses the lancet.  Neither should be  used for more than one person.
       - Whenever possible, blood  glucose meters should not be shared.  If they must be shared, the  device should be cleaned and disinfected after every use, per manufacturer’s  instructions.  If the manufacturer does not specify how the device should  be cleaned and disinfected then it should not be shared. 
          
  
 
        
 
 
 
   
      
 Fingerstick/Lancing  Devices      - My facility uses reusable fingerstick  devices.  If we change the lancet and  disposable components and clean and disinfect the device after use, is it okay  to use this device for multiple patients? Single-use, auto-disabling lancets  are more expensive.
No.  Fingerstick  devices must never be used for more than one person.    Due to failures to change the disposable components,  difficulties with cleaning and disinfection of reusable components after every  use, and their link to multiple hepatitis B virus (HBV) infection outbreaks [3-5],  CDC and FDA recommend that these devices never be used for more than one person [6, 7].  Use of fingerstick devices for more than one person unnecessarily  compromises patient safety, as demonstrated by numerous HBV infection outbreaks  and resulting deaths [4, 8].  Despite  perceived cost-savings from multi-patient use of reusable fingerstick devices,  facilities should also consider the additional costs of testing, treatment, and  legal action that result from such outbreaks and patient notifications.      - Some of the newer fingerstick devices  come with cartridges that have multiple lancets preloaded. Is it acceptable to  use this type of device for multiple patients so long as you remember to  advance to a new lancet each time?
No. These devices are not approved nor  safe for use on multiple patients. Even if the device is advanced and a new  lancet is used for each fingerstick procedure, unused lancets could become  contaminated through contact with blood remaining on the end cap or the device barrel [9].  At least one outbreak of HBV infection  resulting from multi-patient use of these devices has occurred in recent years [9].      - My facility uses reusable fingerstick  devices.  However, we dedicate them for  single-patient use.  Is this acceptable?
CDC recommends the use of single-use, auto-disabling  fingerstick devices in settings where assisted blood glucose monitoring is  performed.  This practice prevents  inadvertent reuse of fingerstick devices for more than one person.  Additionally, the use of single-use,  auto-disabling fingerstick devices protects healthcare personnel from  needlestick injuries.  If reusable fingerstick devices are used for  assisted monitoring of blood glucose then they should be treated in a manner  similar to other personal care items (e.g., razors and toothbrushes) and must  never be shared.  Facilities must take  steps to assure that fingerstick devices are clearly labeled and stored in a  manner to prevent inadvertent use for the wrong patient and cross-contamination  from the surface of one fingerstick device to another (see Question 3 under  Blood Glucose Meters for additional information on storage).      - Residents at our assisted living facility do  their own blood glucose monitoring and prefer to use the reusable fingerstick  devices.  Is this acceptable?     
Reusable fingerstick devices are appropriate for  individuals who perform all steps of testing themselves.  However, this equipment should be labeled  with their name and these individuals should be educated that this equipment  should be treated like other personal care equipment (e.g., razors,  toothbrushes) and must never be shared.   Transmission of HBV infection has been described in residential settings  when individuals shared their personal blood glucose monitoring equipment with  friends or family [10, 11].      
 
 
        
 
 
 
   
      
 Blood  Glucose Meters      - How can hepatitis B virus (HBV) be  transmitted through the meter? If the blood glucose meter never touches the  patient, why does it need to be cleaned and disinfected after each use?
Infectious agents, such as HBV, can be transmitted  through indirect contact transmission, even in the absence of visible blood [4]. Indirect contact transmission is defined as  the transfer of an infectious agent (e.g., HBV) from one patient to another through  a contaminated intermediate object (e.g., blood glucose meter) or person (e.g.,  healthcare personnel hands) [12].  With some blood glucose meters that require pre-loading  of the test strip, the device may come into direct or close contact with the  patient’s fingerstick wound.  If blood is  transferred from the patient to the meter, and the meter is not cleaned and  disinfected after use, subsequent patients can be exposed to this blood when  the meter is used on them.  Indirect contact transmission can also occur even if  the patient never directly contacts the meter. Healthcare personnel hands can become contaminated with blood at various  points while performing assisted blood glucose monitoring including pricking the patient’s finger or handling the test strip.  Blood can then be transferred to the meter  when healthcare personnel handle the meter to obtain the reading.  If the meter is not cleaned and disinfected  after use, the blood remaining on the meter can be transferred to subsequent  patients via healthcare personnel hands when they handle the meter and then  assist with fingerstick procedures.  Numerous  outbreaks have implicated this mechanism in the spread of HBV infections [3, 4].  Contamination of equipment and  transmission of HBV can also occur if healthcare personnel fail to change their  gloves and perform hand hygiene between patients.  A multi-hospital study of blood glucose meters found  that 30% were contaminated with blood; contamination was identified at the test  strip insertion site as well as on the outside surfaces of meters [13].  Further, HBV has been demonstrated to remain  infectious in dried blood on environmental surfaces for at least 7 days [14].   For these reasons, blood glucose meters  should be cleaned and disinfected after each use, unless they are dedicated to  a single patient and appropriately stored to prevent inadvertent contamination  (See Question 3 under Blood Glucose Meters).      - What products are acceptable for cleaning and  disinfection of blood glucose meters? 
FDA has recently released guidance for manufacturers  regarding appropriate products and procedures for cleaning and disinfection of  blood glucose meters.  This guidance, including a link to the Environmental Protection Agency (EPA) website can be  found at FDA's Website.  An excerpt from this guidance reads:  “The  disinfection solvent you choose should be effective against HIV, Hepatitis C,  and Hepatitis B virus. Outbreak episodes have been largely due to transmission  of Hepatitis B and C viruses. However, of the two, Hepatitis B virus is the  most difficult to kill. Please note that 70% ethanol solutions are not  effective against viral bloodborne pathogens and the use of 10% bleach  solutions may lead to physical degradation of your device. View a list of  Environmental Protection Agency (EPA) registered disinfectants effective  against Hepatitis B”  Healthcare  personnel should consult the manufacturers of blood glucose meters in use at  their facilities to determine what products, meeting the criteria specified by  the FDA, are compatible with their meter prior to using any EPA-registered  disinfectant for disinfection purposes.   If manufacturers are unable to provide this information then the meter  should not be used for multiple patients.     - If blood glucose meters are dedicated for  single-patient use, where should they be stored? 
Blood glucose meters dedicated for single-patient  use should, ideally, be stored in the patient’s room in a manner that will  protect against inadvertent use for additional patients and cross-contamination  via contact with other meters or equipment.  An evaluation of instrument storage areas in  hospital found that 20% of areas where blood glucose meters were stored were  contaminated with blood [13].  If  facilities are not able to safely store meters in patient rooms, they need to  take steps to ensure that meters are not inadvertently used for the wrong  patient and that cross-contamination from the surface of one meter to another  does not occur.  If the blood glucose meter becomes contaminated through  inappropriate storage, subsequent patients could be exposed to infectious  agents, even if the meter itself does not have direct patient contact (see  Question 1 under Blood Glucose Meters).     - If blood glucose meters are dedicated for  single-patient use, do they need routine cleaning and disinfection?  If so, how often? 
If meters are dedicated for single-patient use and  facilities have taken steps to assure that they are stored in a location to  prevent inadvertent use for the wrong patient and/or cross-contamination (see  Question 3 under Blood Glucose Meters), then meters should be cleaned and disinfected  according to manufacturer’s instructions and, at a minimum, anytime they are  being reassigned to a different patient.   Facilities are reminded, however, that if the manufacturer of the device  in use does not specify how the device should be cleaned and disinfected, then  it should not be shared or reassigned to a different patient (see Question 1 under General and information at the page on CDC's Injection Safety website titled Infection Prevention during Blood Glucose Monitoring and Insulin Administration).  Care must be taken by personnel handling meters,  whether designated for multi- or single-patient use, to remove gloves and perform  hand hygiene after each patient use and after cleaning and disinfecting meters.       
 
 
        
 
 
 
   
      
 Insulin  Pens and Insulin Administration      - My facility uses insulin pens.  If we change the needle and/or insulin  cartridge, is it okay to use this device for multiple patients? 
No. Insulin  pens are approved and labeled only for single-patient use. Under no circumstances may they be used for  more than one person. Part of safe  injection practices includes never using the same syringe for more than one  patient [12].  Changing only the  needle and reusing the cartridge of an insulin pen is a form of syringe reuse  that represents a serious medical error. Changing the cartridge does not  protect against contamination and does not make these devices safe for  multi-patient use [15].  If insulin pens are in use in a facility, they  should be clearly labeled with the patient’s name and stored in a manner to  prevent inadvertent use for more than one person and/or cross-contamination.  Failure to do so has resulted in large scale patient notifications and an alert  from FDA reminding consumers and healthcare personnel that these devices must  never be used for more than one person [16].     - My facility uses multi-dose vials of insulin.  Can these vials be used for more than one  person?
Multi-dose vials should be dedicated to a single  patient whenever possible.  If they must  be used for more than one person, they should not be stored or accessed in the  immediate patient treatment area.  This  is to prevent inadvertent contamination of the vial through direct or indirect  contact with potentially contaminated surfaces or equipment that could then  lead to infections in subsequent patients.   If a multi-dose vial enters the immediate patient treatment area (e.g.,  patient room), it should be dedicated to that patient only.  Additional information related safe injection  practices and handling of multi-dose vials.      
 
 
        
 
 
 
   
      
 References      - Infection prevention during blood glucose monitoring and insulin administration.
    - Klonoff DC, Perz JF. Assisted monitoring of  blood glucose: Special safety needs for a new paradigm in testing glucose. [PDF - 5 pages] J Diabetes Sci Technol 2010;4(5):1027-1031
    - Centers for Disease Control and Prevention.  Transmission of hepatitis B virus among persons undergoing blood glucose  monitoring in long-term-care facilities-Mississippi, North Carolina, and Los  Angeles County, California, 2003-2004. MMWR 2005;54:220-223.
    - Thompson ND, Perz JF. Eliminating the blood: Ongoing outbreaks of hepatitis B virus infection  and the need for innovative glucose monitoring techniques. [PDF - 6 pages] J Diabetes Sci Technol 2009;3(2):283-288.
    - Polish LB, Shapiro CN, Bauer F, Klotz P, Ginier  P, Roberto RR, Margolis HS, Alter MJ. Nosocomial transmission of hepatitis B  virus associated with a spring-loaded finger-stick device. N Engl J Med  1992;326:721-5.
    - CDC Clinical Reminder: Use of fingerstick devices  on more than one person poses risk for transmitting bloodborne pathogens.
    - FDA Communication: Use of fingerstick devices on  more than one person poses risk for transmitting bloodborne pathogens.
    - Centers for Disease Control and Prevention. Notes  from the field: Deaths from acute hepatitis B virus infection associated with  assisted blood glucose monitoring in an assisted-living facility - North  Carolina, August-October 2010. MMWR 2011;60:182.
    - Gotz HM, Schutten M, Boorsboom GJ, Hendriks B,  van Doornum G, de Zwart O. A cluster of hepatitis B virus infections associated  with incorrect use of a capillary blood sampling device in a nursing home in  the Netherlands, 2007. Eurosurveillance 2008;13:1-5.
    - Farkas K, Jermendy G. Transmission of hepatitis  B infection during home blood glucose monitoring. Diabetic Medicine  1997;14:263.
    - Stapleton J. Transmission of hepatitis B during  blood glucose monitoring. JAMA 1985;253:3250.
    - Centers for Disease Control  and Prevention (CDC). Guideline for isolation precautions:      preventing transmission of  infectious agents in healthcare settings 2007. [PDF - 225 pages] Atlanta (GA): US Department of  Health and Human Services. CDC; 2007.
    - Louie RF, Lau MJ, Lee JH, et al. Multicenter  study of the prevalence of blood contamination on point-of-care glucose meters  and recommendations for controlling contamination. Point of Care  2005;4:158-163.
    - Bond WW, Favero MS, Petersen NJ, et al. Survival  of hepatitis B virus after drying and storage for one week.  Lancet 1981;1(8219):550-1.
    - CDC Injection Safety Frequently Asked Questions
    - FDA: Information  for Healthcare Professionals: Risk of transmission of blood-borne pathogens  from shared use of insulin pens.
    
 
 
 
 
        
 
 